KJ5000 is a recirculating air-cooled water chiller that provides a clean, environmentally friendly and reliable source of temperature-controlled fluid for closed loop systems up to 16L/min. It eliminates scale buildup, sewerage cost, and tap water cooling expenses. This small chiller provides high cooling capacity at low cost and has a convenient digital reading and setting display. Temperature range can be adjusted from 5 to 30°C. It is suitable for MTI RTP Furnaces, High-Temperature Ovens, and any lab equipment that need water cooling.
SPECIFICATIONS
| Power |
- AC 100-120V or AC 208-240V, 50/60Hz
- 0.5 kW
|
| Refrigerant & Cooling Capacity |
- Refrigerant: R134a (~300g)
- Cooling Capacity: 722Kcal/h; 2866Btu/h; 840W
|
| Temperature Control |
|
| Temperature Accuracy |
|
| Max. Pump Head |
|
| Pipe Connectors |
- Φ 12mm PTC connector
- 8mm I.D, 12mm O.D. polyurethane tube is included
|
| Water Flow Rate |
|
|
Water Tank
|
- 6 Liters Water Tank made of stainless steel
- Water chiller should not be connected to a system with the capacity more than the tank's capacity itself.
- Distilled water is required as the coolant.
- Do not use deionized water as the cooling agent, DI water may cause the copper pipe to corrode, mainly pitting, and leak over time.
- Water Drain Plug Threads and Pitch Size: M16, 1.75mm
|
| Water Filter |
- One water filter is included for immediate use.
- Replace the water filter every 3 to 6 months based on water quality and usage. Please click on pic to order
|
| Protection |
- Protected from overflow alarm and over temperature
|
| Noise Level |
- When compressor and water pump is running,
- 70 dB (20 cm away from the chiller)
- 64 dB (100 cm away from the chiller)
- When only the water pump is running,
- 47 dB (20 cm away from the chiller)
- 43 dB (100 cm away from the chiller)
- (Background noise level: 45 dB)
|
| Compliance |
- CE approval
- NRTL or CSA Certification is available upon request at extra cost.
|
| Application Notes |
- Use of tap water is prohibited.
- Please connect the chiller water outlet to the equipment water inlet and the chiller water inlet to the equipment outlet. Otherwise, the water flow sensor on the equipment may fail to work due to the wrong water flow direction
- If you need a temperature below 20C, please set Parameter F9 to the desired temperature(10C-20C).
|